Biological Safety Cabinets

Biological Safety Cabinets are safety equipments to be used for handling Hazardous Pathogens. These cabinets, apart from creating a Clean & Sterile environment, to protect the product/sample also protects the Lab personnel and the Atmosphere. 

The vertical downward airflow protects the product, the inflow along the front edge protects the user and the HEPA filtered exhaust air protects the atmosphere. We are looking for inqueries from South India & Overseas Countries.

Pathogens have been classified into four risk levels from 1 to 4. Based on the level of risk of the pathogen and the procedure, NSF-49 of USA has classified the Biosafety cabinets into three Classes viz. Class I, Class II & Class III. 

Class II cabinets are further classified into 3 types viz. Type A/B3, B1 & B2. 

Class II Type A/B3 recirculates 70% air and exhausts 30% air 
Class II Type B1 recirculates 30% air and exhausts 70% air
Class II Type B2 exhausts 100% air without recirculation.

Biological Safety Cabinet, Class II Type B2

₹ 1.30 Lakh /Unit Get Latest Price

Product Brochure
Lighting TypeLED
Powerless than 1kW
Usage/ApplicationCulture Lab, Microbiology Lab
MaterialGI powder coated & SS304
Cabinet MaterialStainless Steel, Mild Steel
Main FilterHEPA Filter, ULPA filter
SterilizationUV Germicidal Tube
Electric Supply230V, Single Phase
Noise LevelLess than 65 db
Working area size3 ft, 4ft, 6ft Wide X 2ft deep X 2'3" high

Minimum order quantity: 1 Unit

BioMARK-II-B2 is a Class II Type B2 Biohazard Safety Cabinet or Total Exhaust Cabinet. These cabinets provide personnel, product and environment protection. Type B1 cabinets have 0% re-circulation arrangement and 100% exhaust through HEPA filter. Vertical downward airflow to the work zone creates parallel streams of ultra-clean air that carries with it all the contaminants leaving the work area at a high degree of cleanliness of the order of Class 100.

Standard Features :

The filtration arrangement for this Type B1 cabinet is designed as per NSF Standard 49.
All contaminated plenums are surrounded by negative pressure, virtually eliminating leakage from the cabinet. Provides maximum protection.
Blowers are fully balanced and are made for continuous operation. Motor capacity is optimized to deliver the required air with low noise.
The cabinet work zone has minimum welded joints to avoid collection of contaminants or rusting
Cabinet work area is made of fully SS304 construction with minimum joints.
Maximum front sash opening is 8" which can be increased for introducing larger instruments or samples and for cleaning. An alarm is provided to prevent opening of front door beyond 8"
UV germicidal lamp (15W or 30W) is provided for preliminary sterilization..
Built-in white, electronically ballasted, fluorescent lighting provides excellent illumination in the work zone and reduces operator fatigue.
Powder coated GI construction in 2 colour combination provides a pleasing look and enhances the aesthetic appeal of the equipment, Fully SS304 construction is also available.
Exhaust Module with HEPA filter (has to be located near the Cabinet above the false ceiling or outdoors on sunshade etc. with access, for service.)
Every Cabinet is tested in the factory for HEPA filter integrity, Cleanliness and Air flow. Additionally these are repeated at the Site again as per requirement. All Test instruments are Calibrated with traceability to NABL or International Standards.
The cabinets is warranted for 12months excluding consumable parts and accessories.
Additional IQ/OQ documentation is available upon request.

Optional Items: Virus Burnout Unit or Second stage of Exhaust HEPA filter

Note 1: When air is exhaust from the Biosafety Cabinet, not only pathogens but air conditioned air is also thrown out. To compensate for the air lost, air has to be introduced into the room using Positive Pressure / Fresh Air Modules, to avoid leakage into the room through door opeingins / gaps. If fresh air is not fed from an air conditioned room, the lab is heated up to cool this additional Air conditioners have to be installed. As a thumb rule for every 200cfm of fresh air added one ton of AC should be added. For eg. for a 3' Type B2 cabinet, minimum of 2TR AC should be added.

Note 2: For a 3ft Biosafety Cabinet the exhaust air quantity is 450cfm, to heat this air even to to 80degC, 15kW of heater is required, which is too high considering power situation in India. To avoid this we have developed an alternative system that is being used in USA and Europe, which is an additional stage of HEPA filtration in exhaust. An additional HEPA filter Module with a blower and HEPA filter will be connected to the exhaust system. It may be noted that all the HEPA filters in the Biosafety cabinet have to be tested for Integrity using Aerosol Photometer after Installation at Site to ensure safety of the environment.

Bio-Safety Cabinets Class II Type A1

₹ 72,000 /Piece Get Latest Price

Product Brochure
Lighting TypeLED
Usage/ApplicationCulture Lab, Microbiology Lab
MaterialGI POWDER COATED
BrandMARK AIR
Biosafety LevelBSL 3

bioMARK-II-A2 is a Class II Type A2 or A/B3 Biohazard Safety Cabinet. These cabinets provide personnel, product and environment protection. Type A2 cabinets have 70% re-circulation arrangement and 30% exhaust, both through HEPA filters. Vertical downward airflow to the work zone creates parallel streams of ultra-clean air that carries with it all the contaminants leaving the work area at a high degree of cleanliness of the order of Class 100.

Standard Features :

The filtration arrangement for this Type A2 cabinet is designed as per NSF Standard 49.
All contaminated plenums are surrounded by negative pressure, virtually eliminating leakage from the cabinet. Provides maximum protection.
Blowers are fully balanced and are made for continuous operation. Motor capacity is optimized to deliver the required air with low noise.
The cabinet work zone has no welded joints to avoid collection of contaminants or rusting
Cabinet work area is made of fully SS304 construction with minimum
Maximum front sash opening is 8" which can be increased for introducing larger instruments or samples and for cleaning. An alarm is provided to prevent opening of front door beyond 8"
UV light timer is provided to know the usage of UV lamp.
Built-in warm white, electronically ballasted, fluorescent lighting provides excellent illumination in the work zone and reduces operator fatigue.
Powder coated GI construction in 2 colour combination provides a pleasing look and enhances the aesthetic appeal of the equipment.
Every Cabinet is tested in the factory for HEPA filter integrity, Cleanliness and Air flow. These are repeated at the Site again on request. All Test instruments are Calibrated with traceability to NABL or International Standards.
The cabinets is warranted for 12months excluding consumable parts and accessories.
Additional IQ/OQ documentation is available upon request.

Biological Safety Cabinets

₹ 80,000 /Piece Get Latest Price

Product Brochure
Cabinet MaterialMild Steel, Stainless Steel
Main FilterHEPA Filter
SterilizationUV Germicidal Tube
Electric SupplySingle Phase,230V (less than 1kW)
Noise LevelLess than 65 db
BrandbioMARK-Advanced
Size91cm X 62cm X 66cm
Working area size3ft,4ft & 6ft

Minimum order quantity: 1 Piece

Open-fronted equipment with atmost protection for User, Product and the Environment. Ideal for handling Pathogens. HEPA filters / ULPA Filters Tested at factory and onsite* See all variants and choose the most appropriate for your application and risk level. Popular Reliable model, 100s of installations in India.
Warranty and AMC available.

Standard Sizes: Workspace: 29" X 24" X 26", 35" X 24" X 26", 47" X 24" X 26", 59" X 24" X 26", 72" X 24" X 26"

Specifications/Features:

  • State of the Art HEPA / ULPA filters
  • Ergonomic Inclined design for maximum comfort
  • Fully SS304 interiors with powder coated exteriors
  • Removable SS top work table with drain tray
  • Heavy duty Motor-Blower for continuous operation
  • Sliding type front door - single piece with toughened glass with counter weight arrangement with safety alarm
  • Imported Magnehelic Gauge for Filter Pressure drop reading
  • UV Germicidal lamp with hour meter
  • Fluorescent lamp
  • *Performance Test Certificate: HEPA filter integrity; Inflow, downflow exhaust airflow; HEPA filter pressure drop
  • Options: Exhaust Duct, LCD Display alarm for filter usage, Auto-speed control, Microprocessor / Microcontroller based system; Virus Burnout System or additional stage of HEPA filtration
* Onsite Test on request - conditions apply

Performance: Class100 / Passes HEPA filter Integrity Test
Downflow 90 + 20 fpm
Initial Pressure drop 12mm (Max 25mm)

Bio-Safety Cabinets Class II Type B1

₹ 80,000 /Piece(s) Onwards Get Latest Price

BrandMark Air
ClassClass II
Product TypeType B1
Nominal Size3 ft
ModelbioMARK-BI-3
Lighting TypeFluorescent
PowerSingle Phase
Usage/ApplicationDiagnostic Labs, Hospitals & Microbial Research Lab
MaterialMild Steel
Air Flow DirectionVertical
Biosafety LevelBSL 1, BSL 3, BSL 2

Minimum order quantity: 1 Piece(s) Onwards

bioMARK-II-B1 is a Class II Type B1 Biohazard Safety Cabinet. These cabinets provide personnel, product and environment protection. Type B1 cabinets have 30% re-circulation arrangement and 70% exhaust, both through HEPA filters. Vertical downward airflow to the work zone creates parallel streams of ultra-clean air that carries with it all the contaminants leaving the work area at a high degree of cleanliness of the order of Class 100.

Standard Features :

The filtration arrangement for this Type B1 cabinet is designed as per NSF Standard 49.
All contaminated plenums are surrounded by negative pressure, virtually eliminating leakage from the cabinet. Provides maximum protection.
Blowers are fully balanced and are made for continuous operation. Motor capacity is optimized to deliver the required air with low noise.
The cabinet work zone has no welded joints to avoid collection of contaminants or rusting
Cabinet work area is made of fully SS304 construction with minimum
Maximum front sash opening is 8" which can be increased for introducing larger instruments or samples and for cleaning. An alarm is provided to prevent opening of front door beyond 8"
UV light timer is provided to know the usage of UV lamp.
Built-in warm white, electronically ballasted, fluorescent lighting provides excellent illumination in the work zone and reduces operator fatigue.
Powder coated GI construction in 2 colour combination provides a pleasing look and enhances the aesthetic appeal of the equipment.
Every Cabinet is tested in the factory for HEPA filter integrity, Cleanliness and Air flow. These are repeated at the Site again on request. All Test instruments are Calibrated with traceability to NABL or International Standards.
The cabinets is warranted for 12months excluding consumable parts and accessories.
Additional IQ/OQ documentation is available upon request.

Class III Bio-Safety Cabinets

₹ 1 Lakh /Piece(s) Onwards Get Latest Price

BrandMark Air
ModelbioMARK-III-3
Nominal Size3 ft
ClassClass III
Usage/ApplicationDiagnostic Labs, Hospitals & Microbial Research Lab
MaterialGI/MS & SS
Air Flow DirectionVertical
Biosafety LevelBSL 2, BSL 3, BSL 4, BSL 1
Inflow VelocityN A
Size3 X 2 X 2ft, 4 X 2 X 2ft & 6 x 2 X 2ft
FeaturesGI powder coated with SS workspace. Made & Tested at Site as per NSF-49
ColorOff-white & Blue

Minimum order quantity: 1 Piece(s) Onwards

Recommended for Product, Personnel & Environment protection only Air Balancing: 100% exhaust with incineration & working zone under negative pressure. As the power consumption for incineration is high (around 10kW), we suggest 2-stage exhaust filtration through HEPA/ULPA.

bioMARK-III, Class III Biological Safety Cabinet is designed for work with Microbiological agents assigned to biosafety level 4, and provides maximum protection to the environment and the worker. Access for passage of materials into the cabinet is through a double-door pass-through box that can be decontaminated between uses. Both supply and exhaust air are HEPA filtered. The exhaust air passes through two stage HEPA Filters, or a HEPA filtered and an air incinerator, before discharge to the outdoors. Airflow is maintained by a dedicated independent exhaust blower outside the cabinet, which keeps the cabinet safe, i.e. under negative pressure.
Arm-length, heavy-duty rubber gloves are attached in a gas-tight manner to ports in the cabinet and allow for manipulation of the materials isolated inside. Although these gloves restrict movement, they prevent the users direct contact with the hazardous materials.

Standard Features :

  • Exhaust air is double-filtered through high-quality HEPA/ULPA filters (per IEST-RP-CC-001.3) with typical efficiency of >99.999% for 0.1 to 0.3 micron particles
  • Exclusive dual exhaust filters provide >100,000 times better protection than single-stage designs.
  • Magnehelic pressure gauge* is mounted in the rear of the work zone for at-a-glance monitoring of work zone negative pressure.
  • Neoprene gauntlets are single piece leak-tested glove assemblies which guarantee maximum protection.
  • An integrated pass-through with interlocking doors, permits materials transfer without risk of environmental contamination.
  • Available in GI powdercoated or SS304 construction with leak free welded-sealed joints.
  • An angled cabinet front ensures an ergonomic working posture.
  • Cabinet operates at negative pressure relative to the laboratory in order to prevent migration of pathogenic materials out of the work area.
  • All cabinets individually tested for important paramenters as per NSF-49 with capability to Calibrate on Site after installation or periodically.
  • Calibrated Test Instruments used with traceability to NABL or (Inter)National Standards.

Biosafety Cabinet Manufacturer

₹ 95,000 /Piece Get Latest Price

Lighting TypeLED
Usage/ApplicationLABORATORY
MaterialStainless Steel
BrandSTANDMARK
Biosafety LevelBSL 3

Mark Air Class I biohazard safety cabinet is maintained under negative pressure, ventilated cabinet usually operated with an open front. All of the air from the cabinet is exhausted through a HEPA either into the laboratory or to the out side. The Class I BSC is designed for general microbiological research with low-and moderate-risk agents, and is useful for containment of mixers, blenders, and other equipment. These cabinets are not appropriate for handling research materials that are vulnerable to airborne contamination, since the inward flow of unfiltered air from the laboratory can carry microbial contaminants into the cabinet.

Bio-Safety Cabinets Class II Type A2

₹ 72,000 /Piece(s) Onwards Get Latest Price

ModelbioMARK-A2-3
Nominal Size3 ft
Product TypeType A2
ClassClass II
Usage/ApplicationDiagnostic Labs, Hospitals & Microbial Research Lab
MaterialSS & MS
Air Flow DirectionVertical
Biosafety LevelBSL 2, BSL 3, BSL 1
Inflow Velocity90 to 100fpm
Voltage230V
Size3 X 2 X 2ft, 4 X 2 X 2ft & 6 x 2 X 2ft
PhaseSingle
FeaturesGI powder coated with SS workspace. Made & Tested at Site as per NSF-49
ColorOff-white & Blue

Minimum order quantity: 1 Piece(s) Onwards

bioMARK-II-A2 is a Class II Type A2 or A/B3 Biohazard Safety Cabinet. These cabinets provide personnel, product and environment protection. Type A2 cabinets have 70% re-circulation arrangement and 30% exhaust, both through HEPA filters. Vertical downward airflow to the work zone creates parallel streams of ultra-clean air that carries with it all the contaminants leaving the work area at a high degree of cleanliness of the order of Class 100.

Standard Features :

The filtration arrangement for this Type A2 cabinet is designed as per NSF Standard 49.
All contaminated plenums are surrounded by negative pressure, virtually eliminating leakage from the cabinet. Provides maximum protection.
Blowers are fully balanced and are made for continuous operation. Motor capacity is optimized to deliver the required air with low noise.
The cabinet work zone has no welded joints to avoid collection of contaminants or rusting
Cabinet work area is made of fully SS304 construction with minimum
Maximum front sash opening is 8" which can be increased for introducing larger instruments or samples and for cleaning. An alarm is provided to prevent opening of front door beyond 8"
UV light timer is provided to know the usage of UV lamp.
Built-in warm white, electronically ballasted, fluorescent lighting provides excellent illumination in the work zone and reduces operator fatigue.
Powder coated GI construction in 2 colour combination provides a pleasing look and enhances the aesthetic appeal of the equipment.
Every Cabinet is tested in the factory for HEPA filter integrity, Cleanliness and Air flow. These are repeated at the Site again on request. All Test instruments are Calibrated with traceability to NABL or International Standards.
The cabinets is warranted for 12months excluding consumable parts and accessories.
Additional IQ/OQ documentation is available upon request.

Virus Burn Out Unit

₹ 35,000 /Piece Get Latest Price

BrandMark Air
Product TypeType B2
Nominal SizeAll size
ModelMARK-B2-VIR-4
Power12 kW
Usage/ApplicationExhaust Air treated to 120 deg C using electrical strip heaters (3 Phase)
MaterialStainless Steel
Air Flow DirectionHorizontal

Fabricated by using IS 304 stainless steel, equipped with high resistance electric heater with EPS / glass wool insulation, automatic temperature controlling thermostat to raise the exhaust air temperature to 80-degree Celsius (Adjustable), to incinerate / destroy hazardous pathogens in the exhaust air to protect the environment.

Suits to standard size of Bio-Safety Cabinets:

Direction of Flow

Vertical Flow (100% Air Exhaust) Suitable for Biological Safety Cabinet (Type B2)
Capacity (Airflow)450cfm600cfm

900cfm

Suitable to BSC

W 3 x D 2 x H 2

W 4 x D 2 x H 2

W 6 x D 2 x H 2

Virus Burn-Out Size

200 x 200 x 600 mm

250 x 250 x 600 mm

300 x 300 x 650 mm


Note 1: When air is exhaust from the Biosafety Cabinet, not only pathogens but air conditioned air is also thrown out. To compensate for the air lost, air has to be introduced into the room using Positive Pressure / Fresh Air Modules, to avoid leakage into the room through door opeingins / gaps. If fresh air is not fed from an air conditioned room, the lab is heated up to cool this additional Air conditioners have to be installed. As a thumb rule for every 200cfm of fresh air added one ton of AC should be added. For eg. for a 3' Type B2 cabinet, minimum of 2TR AC should be added.

Note 2: For the 3ft Biosafety Cabinet the exhaust air quantity is 450cfm, to heat this air to 80degC, 15kW of heater is required, which is too high considering power situation in India. To avoid this we have developed an alternative system that is being used in USA and Europe, which is an additional stage of HEPA filtration in exhaust. An additional HEPA filter Module with a blower and HEPA filter will be connected to the exhaust system. It may be noted that all the HEPA filters in the Biosafety cabinet have to be tested for Integrity using Aerosol Photometer after Installation at Site to ensure safety of the environment.

Class I Bio-Safety Cabinets

₹ 1.10 Lakh /Piece Get Latest Price

Product Brochure

Mark Air Class I biohazard safety cabinet is maintained under negative pressure, ventilated cabinet usually operated with an open front. All of the air from the cabinet is exhausted through a HEPA either into the laboratory or to the out side. The Class I BSC is designed for general microbiological research with low-and moderate-risk agents, and is useful for containment of mixers, blenders, and other equipment. These cabinets are not appropriate for handling research materials that are vulnerable to airborne contamination, since the inward flow of unfiltered air from the laboratory can carry microbial contaminants into the cabinet.

X

Contact Us

M. Raghu (Manager)
Mark Air Particulate Control Systems
New No. 5A, Old No. 80, 53rd Street, Opposite Anjaneyar Temple, Ashok Nagar
Chennai - 600083, Tamil Nadu, India

Get Directions
Send Email
Share:
Thank youYour Enquiry has been sent successfully.